Ethylene Oxide Sterilization Operation Procedure


(1) According to the test requirements, prepare the spore suspension and bacterial slices of Bacillus subtilis var. subtilis var. niger. The bacterial slices are sealed in double-layer polyethylene plastic bags, 2 slices per bag. 20 bags are used for each test.
(2) Set one point in the inner, middle and outer layers of the sterilizer, for a total of 9 points. Place one bag in the middle of each point, for a total of 18 bags of test bacterial slices. Place another bag (2 slices) of bacterial slices at room temperature as a positive control.
(3) Perform ethylene oxide sterilization under full load conditions according to the ethylene oxide concentration, action time, temperature and relative humidity in the cabinet specified in the instruction manual. The full load depends on the purpose of the sterilizer. After the treatment, take out the bacterial slices, transfer them to 5 ml nutrient broth culture medium, and culture them in a 37℃ incubator. Observe the final results after 7 days (qualitative culture test).
For nutrient broth tubes with difficult-to-judge results, take 0.2 ml of the suspension and inoculate the nutrient agar plate. Use a sterile L stick to spread evenly and place in a 37°C incubator for culture. After 48 hours, stain the smear and observe the colony morphology under a microscope, or perform other tests to determine whether there is growth or whether the growth is the test bacteria. If it is not the test bacteria, the test should be repeated.
(4) During the test, qualitative and quantitative positive control groups and negative control groups should be set up at the same time.
(5) For the quantitative positive control group, place the same batch of test bacterial slices at room temperature. After the disinfection or sterilization test group reaches the specified action time, immediately place two pieces of the bacterial slices in test tubes containing 5.0 ml PBS and shake each for 80 times. Take the wash solution and count the live bacteria culture according to the method shown in 2.1.1.3.
(6) For the qualitative positive control group, place the test bacterial slices from the same batch at room temperature. After the sterilization test group reaches the specified action time, immediately inoculate 2 of the bacterial slices into 5.0 ml of nutrient broth medium, place them in an incubator for qualitative culture, and observe whether there is bacterial growth.
(7) For the negative control group, inoculate 2 uncontaminated sample slices from the same batch into 5.0 ml of nutrient broth medium, and place the uninoculated nutrient broth medium in an incubator for qualitative culture to observe whether there is bacterial growth.
(8) The test is repeated 5 times.
(9) In each of the 5 tests, the recovered bacterial count of the positive control bacterial slices should be between 5×10 cfu/slice and 5×10 cfu/slice; the qualitative positive control group should have good bacterial growth; and the negative control should have no bacterial growth. If the positive or negative control results do not meet the above requirements, the test will be invalid and repeated.
Evaluation regulations
In the five sterilization tests, the quantitative positive control group in each test had a detected and recovered bacterial count of 5 × 10 cfu/piece to 5 × 10 cfu/piece; the qualitative positive control group had good bacterial growth. The negative control should have no bacterial growth. When all the test bacterial pieces have no bacterial growth, the sterilization is considered qualified.
Precautions
(1) Temperature and relative humidity have a great influence on the sterilization effect of ethylene oxide gas, so the relevant conditions in the test should be strictly controlled.
(2) Ethylene oxide liquid can dissolve polyethylene, polyvinyl chloride, etc., and its liquid should not be dripped on such items. Ethylene oxide, whether liquid or gas, can damage celluloid products, so care should be taken during the test.
(3) Ethylene oxide is flammable and explosive. Fire and explosion prevention measures should be taken at the operation site, and open flames and electric sparks should not be allowed.
(4) Inhalation of excessive ethylene oxide gas can cause poisoning symptoms such as headache and vomiting, and in severe cases, it can cause pulmonary edema. The working environment should have good ventilation. During the 8-hour work day, the ethylene oxide concentration should not exceed 1.82 mg/m3 (1 ppm), and the exposure concentration during the 15-minute work should not exceed 9.1 mg/m3 (5.0 ppm). If poisoning symptoms occur, leave the scene immediately. For mild cases, breathe fresh air until the symptoms disappear; for severe cases, send to hospital for treatment in time.